rule
Rheodyne - A Unit of IDEX Corporation Chromatography ForumSearchContact Us
rule
rule
rule
HomeFluidic ProductsOEM SolutionsTechnical SupportCareersAbout Us
rule
rule
rule
Art
Tech TipsWaveyWaveyWavey
And More...

For more technical support information, also see the Troubleshooting Guide and our on-line product literature:

If you don't find the information you need, use the Fluidic Products Support form to contact our technical support team.

6: Sample Loop Loading

Partial-Filling
Use the partial-filling method if you need to conserve sample or if you want to vary sample volume frequently. In partial-filling, the syringe sets the volume injected onto the column. There is no sample waste, and the volume injected onto the column is equal to that dispensed from the syringe. Reproducibility is 1.0% relative standard deviation (RSD). The volume of the sample loaded is limited to half the sample loop volume. For example, the most you can load into a 200 µL sample loop is 100 µL. See Figure 1. This limitation is due to the manner in which fluids move in tubes. Fluidic movement in tubes affects reproducibility. See Tech Tip 7.

Sample Loop: Partial-Filling

Figure 1. The sample loop can fill up to half the loaded volume in the partial-filling method.

Complete-Filling
Use the complete-filling method if you have a sufficient amount of sample with which to work, if you do not vary sample volume, or if you need high reproducibility. In complete-filling, the loop sets the volume loaded onto the column. You use excess sample (two to five loop volumes) to replace all the mobile phase in the loop. See Figure 2. Change the loop to vary the sample volume. Reproducibility is typically 0.1% RSD for loop sizes > 5 µL. Accuracy is limited as loop volumes are nominal.

Sample Loop: Complete Filling

Figure 2. The sample loop is filled in excess in complete-filling method.

Q: "Which method should I use and which Rheodyne sample injectors use this method?"

A: There are two types of injectors available: dual mode and single mode. Dual mode injectors allow both partial- and complete-filling, whereas single mode injectors allow only complete-filling. See Sample Injectors.

If you are collecting experimental data, sample is scarce, and/or you want to use different sample volumes, a dual mode injector with a large volume sample loop is appropriate. Only dual mode injectors allow the partial-filling method with which you can easily vary your volumes (up to half your sample loop volume) by setting the syringe volume. Once you begin routine analysis and/or you have an abundance of sample, either a dual mode or single mode injector is appropriate. Both types of injectors allow the complete-filling method with which you fill the sample loop in excess. Complete filling maximizes the reproducibility of your results.

Back to Top

spacer
rule
IDEX
rule
rule