
Partial-Filling
Use the partial-filling method if you need to conserve sample
or if you want to vary sample volume frequently. In partial-filling,
the syringe sets the volume injected onto the column. There
is no sample waste, and the volume injected onto the column
is equal to that dispensed from the syringe. Reproducibility
is 1.0% relative standard deviation (RSD). The volume of the
sample loaded is limited to half the sample loop volume. For
example, the most you can load into a 200 µL sample
loop is 100 µL. See Figure 1. This
limitation is due to the manner in which fluids move in tubes.
Fluidic movement in tubes affects reproducibility. See Tech
Tip 7.

Figure 1. The sample loop can fill up to half the loaded volume
in the partial-filling method.
Complete-Filling
Use the complete-filling method if you have a sufficient amount
of sample with which to work, if you do not vary sample volume,
or if you need high reproducibility. In complete-filling,
the loop sets the volume loaded onto the column. You use excess
sample (two to five loop volumes) to replace all the mobile
phase in the loop. See Figure 2. Change the
loop to vary the sample volume. Reproducibility is typically
0.1% RSD for loop sizes > 5 µL. Accuracy is limited
as loop volumes are nominal.
Figure 2. The sample loop is filled in excess in complete-filling
method.
Q: "Which method should I use and which Rheodyne
sample injectors use this method?"
A: There are two types of injectors available:
dual mode and single mode. Dual mode injectors allow both
partial- and complete-filling, whereas single mode injectors
allow only complete-filling. See Sample
Injectors.
If you are collecting experimental data, sample is scarce,
and/or you want to use different sample volumes, a dual mode
injector with a large volume sample loop is appropriate. Only
dual mode injectors allow the partial-filling method with
which you can easily vary your volumes (up to half your sample
loop volume) by setting the syringe volume. Once you begin
routine analysis and/or you have an abundance of sample, either
a dual mode or single mode injector is appropriate. Both types
of injectors allow the complete-filling method with which
you fill the sample loop in excess. Complete filling maximizes
the reproducibility of your results.

|