
Air in the sample loop can cause instantaneous system pressure
drop that eventually returns to a normal level. Air causes
the pressure to drop when the injector moves from the LOAD
to the INJECT position. When large sample loops (>100 µL)
are partially loaded, air present in the needle port tube
is pushed into the sample loop (see Figure
1). Air can also enter the sample loop from siphoning,
which occurs when the vent line is higher than the injection
port. In either case, upon injection, the system pressure
collapses the air bubble, causing pressure to drop momentarily.
A pressure drop in the system caused by air results in changes
in retention time, artifact peaks, and affects column performance.

Figure. 1. Air (white) present in the needle port tube is
pushed by the syringe during loading into the sample loop.
Pressure drops can be avoided by removing the air in the
needle port tube. Do this by flushing about 1 mL of mobile
phase with a luer syringe with needle port cleaner. Keep the
needle port tube filled with mobile phase by occasional flushing.
Adjust the vent line(s) so that the outlet is at the same
horizontal level as the needle port (see Figure
2). For additional injection troubleshooting,
refer to the Rheodyne Troubleshooting
Guide.

Figure 2. Pathway of the flushing mobile phase using the needle
port cleaner when the injector is in INJECT.

|